Resumé
CT-scanninger indebærer ioniserende stråling, så dosis bør holdes lav uden at gå på kompromis med diagnostisk billedkvalitet. I Siemens’ dosismodulering CARE Dose 4D indgår indstillingen organ characteristic, som afspejler forventede attenueringer i forskellige anatomiske områder; der findes dog ingen specifik indstilling til helkrops-CT. Formålet var at undersøge, hvilken organ characteristic (thorax, abdomen, pelvis eller spine) der giver det bedste kompromis mellem dosis og billedkvalitet ved helkrops-CT. I en fantomundersøgelse (CT Whole Body Phantom PBU-60) på en Siemens Biograph Vision 600 Edge blev der udført høj- og lavdosis helkrops-CT med varierende lejringer og størrelser. Dosis blev vurderet med CTDI, objektiv billedkvalitet med SNR og CNR, og subjektiv billedkvalitet med visual grading analysis; forskelle blev testet med One-way ANOVA og Kruskal–Wallis. Dosis var højest med organ characteristic thorax og lavest med pelvis, men der var ingen statistisk signifikant forskel i dosis eller billedkvalitet mellem de fire indstillinger. Konkordansen mellem objektive og subjektive vurderinger var 6,25% ved høj dosis og 31,25% ved lav dosis. Scatterplots tydede på, at thorax gav højeste dosis og bedste billedkvalitet ved høj dosis, mens lavdosis-scanninger havde bedst objektiv kvalitet med thorax og bedst subjektiv kvalitet med pelvis. Samlet peger resultaterne på, at organ characteristic pelvis kan overvejes til helkrops-CT, da den gav den laveste dosis uden påvist forringelse af billedkvaliteten.
CT imaging uses ionizing radiation, so patient dose should be kept low without compromising diagnostic image quality. Siemens’ dose modulation CARE Dose 4D includes an organ characteristic setting that reflects expected attenuation in different anatomical regions, but no setting is dedicated to whole-body CT. This study investigated which organ characteristic (thorax, abdomen, pelvis, or spine) offers the best compromise between dose and image quality for whole-body CT. In a phantom study (CT Whole Body Phantom PBU-60) on a Siemens Biograph Vision 600 Edge, high- and low-dose whole-body scans were acquired with varied positions and sizes. Dose was evaluated using CTDI, objective image quality using SNR and CNR, and subjective image quality using visual grading analysis; differences were tested with One-way ANOVA and Kruskal–Wallis. Dose was highest with thorax and lowest with pelvis, but there were no statistically significant differences in dose or image quality among the four settings. Concordance between objective and subjective assessments was 6.25% for high-dose and 31.25% for low-dose scans. Scatterplots suggested that thorax yielded the highest dose and best image quality at high dose, while at low dose thorax gave the best objective quality and pelvis the best subjective quality. Overall, the findings indicate that the pelvis organ characteristic may be considered for whole-body CT, as it provided the lowest dose without a demonstrated loss of image quality.
[Dette resumé er genereret med hjælp fra AI direkte fra projektet (PDF)]