Termografiske målinger med PX1. Inter- og intrarater-reliabilitet og betydningen af kamera afstand.
Oversat titel
Thermographic Measurements with PX1. Inter- and Intrarater- reliability and the Impact of camera-to-skin distance.
Forfattere
Laursen, Taziya Aggerholm ; Anantharajah, Sowmiya ; Poulsen, Louise Hwiid
Semester
4. semester
Udgivelsesår
2026
Afleveret
2026-05-31
Antal sider
83
Resumé
Denne kandidatafhandling undersøger, om smartphonebaseret infrarød termografi kan bruges pålideligt til at monitorere indstikssteder (pins) hos patienter i ekstern fiksation, hvor kliniske vurderinger ofte er subjektive. Formålet var at vurdere inter- og intrarater-reliabiliteten af PX1, en smartphone med integreret termografisk kamera, samt at afdække betydningen af kamera-hud-afstand. Studiet var et kvantitativt metodologisk design med to delstudier: I delstudie 1 blev patienter med ekstern fiksation termograferet, og temperaturparametrene Max (maksimal temperatur) og T50 (gennemsnittet af de varmeste 50 % inden for ROI) blev anvendt til at vurdere reliabilitet. I delstudie 2 blev raske deltagere termograferet ved 40 cm og 60 cm for at teste afstandens effekt. Resultaterne viste, at intrarater-reliabiliteten var højere end interrater-reliabiliteten for begge parametre; intrarater ICC lå mellem 0,52 og 0,71 (moderat), mens interrater ICC lå mellem 0,39 og 0,53 (lav til moderat). Aftaleanalyser viste bredere grænser for interrater end intrarater sammenligninger. Der var signifikante forskelle mellem målinger ved 40 cm og 60 cm for både T50 og Max (p < 0,001), hvor målinger ved 60 cm var lavere end ved 40 cm. Samlet peger resultaterne på, at PX1-termografi har moderat pålidelighed ved gentagne målinger fra samme rater, men begrænset overensstemmelse mellem forskellige ratere, og at standardiserede måleprocedurer, herunder fastlagt kameraafstand, er afgørende. Smartphonebaseret termografi kan dermed have potentiale som monitoreringsmetode for pins, forudsat ensartet udførelse.
This master’s thesis evaluates whether smartphone-based infrared thermography can be used reliably to monitor pin-sites in patients with external fixation, where clinical assessments are often subjective. The aim was to assess the inter- and intrarater reliability of PX1, a smartphone with an integrated thermal camera, and to examine the influence of camera-to-skin distance. The study used a quantitative methodological design with two sub-studies: In sub-study 1, patients with external fixation were imaged, and the temperature metrics Max (maximum temperature) and T50 (the mean of the warmest 50% within the ROI) were used to evaluate reliability. In sub-study 2, healthy participants were imaged at 40 cm and 60 cm to test the effect of distance. Intrarater reliability exceeded interrater reliability for both metrics; intrarater ICC values ranged from 0.52 to 0.71 (moderate), while interrater ICC values ranged from 0.39 to 0.53 (low to moderate). Agreement analyses showed wider limits for interrater than intrarater comparisons. There were significant differences between measurements at 40 cm and 60 cm for both T50 and Max (p < 0.001), with lower temperatures at 60 cm. Overall, PX1 thermography shows moderate consistency for repeated measurements by the same rater but limited agreement between different raters, and standardized measurement procedures—including fixed camera distance—are essential. Smartphone thermography therefore has potential as a pin-site monitoring method if applied consistently.
[Dette resumé er genereret med hjælp fra AI direkte fra projektet fuldtekst]
Emneord